FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics, REF: RFIT-ASY-0126, and RFIT-ASY-0127, when used...
BioFire Diagnostics, LLC began a recall of FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics, REF: RFIT-ASY-0126, and RFIT-ASY-0127, when used with the following blood culture bottles: Bottle... on . FDA classified it as Class II and gives the reason as "Increased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1133-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- BioFire Diagnostics, LLC
- Product
- FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics, REF: RFIT-ASY-0126, and RFIT-ASY-0127, when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/410851; BACT/ALERT FN Plus/410852; BACT/ALERT PF Plus/410853
- Product code
- PEN Gram-Negative Bacteria And Associated Resistance Markers
- Reason for recall
- Increased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- On 03/07/19, Urgent: Medical Device Recall (Correction) notices were mailed to customers. Customers were informed that the false positive results are primarily Enterobacteriaceae results with a lower rate of concurrent E. coli results. If the affected BCID panel is used with BACT/ALERT blood culture bottles from the lots listed, then positive results for Escherichia coli and Enterobacteriaceae (if occurring without a specific genus/species reported other than E. coli; i.e. negative for all of the following: Enterobacter cloacae complex, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens) should be confirmed by another method prior to reporting the test results. Confirmation is not required if the Enterobacteriaceae detected result is accompanied by a detected result for Enterobacter cloacae complex, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, or Serratia marcescens. Customers were asked to complete and return Acknowledgement of Receipt forms.
- Quantity in commerce
- 20350
- Distribution
- MA, NH, VT, CT, NJ, NY, PA, DE, DC, VA, MD, WV, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, ND, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, WY, ID, UT, AZ, NM, NV, CA, HI, OR, WA, AK, BM
- Product codes and lots
- All lots of BCID Panels, when used with the following blood culture bottles, Part Number/Lot Numbers:410851/4052663, 4052639, 4052640, 4052599, 4052598, 4052547, 4052546, 4052471, 4052472, 4052799, 4052800, 4052815, 4052816; 410852/4052871, 4052872, 4052459, 4052458, 4052565, 4052564, 4052705, 4052704, 4052735, 4052617, 405261; 410853/4052717, 4052718, 4052503, 4052504
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1133-2019 | Class II | Terminated | Voluntary: Firm initiated |