VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658
Ortho-Clinical Diagnostics began a recall of VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in... on . FDA classified it as Class II and gives the reason as "One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1222-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.
- Quantity in commerce
- 80
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom
- Product codes and lots
- Systems Running Software Version 3.4 & 3.4.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1222-2019 | Class II | Terminated | Voluntary: Firm initiated |