Z-1222-2019Class II

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658

Ortho-Clinical Diagnostics, Inc. / initiated 2019-03-11 / Terminated / root cause: software

Ortho-Clinical Diagnostics began a recall of VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in... on . FDA classified it as Class II and gives the reason as "One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1222-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.
Root cause tag
software
FDA root cause
Software design
Action
Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.
Quantity in commerce
80
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom
Product codes and lots
Systems Running Software Version 3.4 & 3.4.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1222-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1222-2019Class IITerminatedVoluntary: Firm initiated