SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung...
LivaNova USA Inc. began a recall of SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a... on . FDA classified it as Class II and gives the reason as "For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1099-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.
- Product code
- DWC Controller, Temperature, Cardiopulmonary Bypass
- Reason for recall
- For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- The recalling firm, LivaLova, contacted the first customer by phone on 9/14/2018 followed by an "IMPORT" letter, the first of which was dated 9/18/2018. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: keep these devices out of service until the investigation is complete. To return your device, please email [email protected]. If taking Serial Numbers 16S15335 and 16S15337 out of service will negatively impact your clinical practice, we can arrange for delivery of loaner devices during this investigation and,as necessary, while they undergo repeat Deep Cleaning Process. If you have any questions, contact V.P., Clinical, Quality, and Regulatory Services at 303-467-6306 or email: [email protected].
- Quantity in commerce
- 6 devices
- Distribution
- US Distribution to states of: IN, MO, and OH.
- Product codes and lots
- Serial numbers 16S13187, 16S14887, 16S15335, 16S15336, 16S15337, and 16S11877.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1099-2019 | Class II | Terminated | Voluntary: Firm initiated |