Z-1099-2019Class II

SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung...

LivaNova USA, Inc. / initiated 2018-09-14 / Terminated / root cause: manufacturing process

LivaNova USA Inc. began a recall of SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a... on . FDA classified it as Class II and gives the reason as "For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1099-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
The recalling firm, LivaLova, contacted the first customer by phone on 9/14/2018 followed by an "IMPORT" letter, the first of which was dated 9/18/2018. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: keep these devices out of service until the investigation is complete. To return your device, please email [email protected]. If taking Serial Numbers 16S15335 and 16S15337 out of service will negatively impact your clinical practice, we can arrange for delivery of loaner devices during this investigation and,as necessary, while they undergo repeat Deep Cleaning Process. If you have any questions, contact V.P., Clinical, Quality, and Regulatory Services at 303-467-6306 or email: [email protected].
Quantity in commerce
6 devices
Distribution
US Distribution to states of: IN, MO, and OH.
Product codes and lots
Serial numbers 16S13187, 16S14887, 16S15335, 16S15336, 16S15337, and 16S11877.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1099-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1099-2019Class IITerminatedVoluntary: Firm initiated