Paltop 1.25 Hex Drivers, Long Part Number: 60-70101
Keystone Dental Inc began a recall of Paltop 1.25 Hex Drivers, Long Part Number: 60-70101 on . FDA classified it as Class II and gives the reason as "Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1063-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Keystone Dental Inc.
- Product
- Paltop 1.25 Hex Drivers, Long Part Number: 60-70101
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Keystone Dental initiated recall by phone to each one of the impacted customers. The phone call was then followed by a Recall Notification Letter that was mailed via overnight FedEx mail on February 22, 2019. Letters states problem and action to take: Replacement 1.25mm Hex Driver(s) have been provided to you with this letter. The 1.25mm Hex Driver(s) currently in your kit should be discarded. Please refer to the images on page 2 to locate the driver within your kit. This issue is limited to this 1.25mm Hex Driver(s) only and no other product within the kit is affected by this issue. Complete the Reply Communication form.
- Quantity in commerce
- 3 hex Driver
- Distribution
- FL, MA, SC, VA
- Product codes and lots
- Lot Number: WO-010352
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1061-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1062-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1063-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1064-2019 | Class II | Terminated | Voluntary: Firm initiated |