Beaver Xstar Slit Knife 2.4 mm, 45degree Double Bevel, Ophthalmic-utilized in Cataract and Glaucoma...
Beaver Visitec began a recall of Beaver Xstar Slit Knife 2.4 mm, 45degree Double Bevel, Ophthalmic-utilized in Cataract and Glaucoma procedures REF # 370824 Product Usage: Ophthalmic... on . FDA classified it as Class II and gives the reason as "Mislabeled. The package labeling indicates a size 2.4mm, 45 degree Double Bevel knife however, the marking on device shows a 2.75mm size device". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1111-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Beaver Xstar Slit Knife 2.4 mm, 45degree Double Bevel, Ophthalmic-utilized in Cataract and Glaucoma procedures REF # 370824 Product Usage: Ophthalmic techniques-utilized in Cataract and Glaucoma procedures
- Product code
- HNN Knife, Ophthalmic
- Reason for recall
- Mislabeled. The package labeling indicates a size 2.4mm, 45 degree Double Bevel knife however, the marking on device shows a 2.75mm size device
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Beaver-Visitec International, Inc.(BVI) initiated recall by letter on 3/12/19 advising users of the problem, action to take: immediately examine inventory and quarantine product from this lot subject to mplete and return the enclosed Product Recall Notice Response & Return Form recall, Return ALL quarantined product from the affected lot. Questions you may contact our Customer Service Team at:[email protected] or call 866-906-8080 from 8 A.M. to 6 P.M. EST (Monday-Friday).
- Quantity in commerce
- 196 boxes (10units/box)
- Distribution
- Worldwide Distribution - US Nationwide Foreign:UNITED KINGDOM JAPAN NETHERLANDS
- Product codes and lots
- Lot Number: 6008554
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1111-2019 | Class II | Terminated | Voluntary: Firm initiated |