Z-1109-2019Class II

Elekta Unity, Image-Guided Radiation Therapy System

Elekta Limited / initiated 2019-03-12 / Terminated / root cause: software

Elekta Limited began a recall of Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments... on . FDA classified it as Class II and gives the reason as "Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1109-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
Root cause tag
software
FDA root cause
Software design
Action
The firm issued a field safety notice on 03/12/2019 by email. The notice explained the issue and cautioned the consignee to select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
Quantity in commerce
12 units
Distribution
Worldwide Distribution - US Nationwide in the states of TX, NY, WI, and Denmark, Germany, Italy, Netherlands, Sweden, United Kingdom
Product codes and lots
Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1109-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1109-2019Class IITerminatedVoluntary: Firm initiated