cobas e 801 immunoassay analyzer
Roche Diagnostics GmbH / initiated 2019-03-15 / Terminated / root cause: design
Roche Diagnostics Corporation began a recall of cobas e 801 immunoassay analyzer on . FDA classified it as Class II and gives the reason as "The firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1249-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- cobas e 801 immunoassay analyzer
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On March 15, 2019, the firm notified affected customers of the recall via an Urgent Medical Device Correction letter. Customers were informed of the potential problem of signal drops when the standby bottle flowpath is covered with a biofilm. Customers were asked to do the following: " Follow the Workaround section to determine if your cobas e 801 module is potentially affected and perform the steps if applicable. " If your cobas e 801 module is potentially affected based on the information in the Workaround section of this UMDC, contact the Roche Support Network Customer Support Center at 1-800-428-2336 to schedule the ProCell II M flowpath decontamination procedure. The ProCell II M flowpath decontamination procedure should be performed every 4 weeks until your cobas e 801 module is switched to the improved ProCell II M formulation. " Identify and retest potentially affected patient samples using the Retrospectively Identify and Correct Potentially Affected Patient Sample Results section of this UMDC. " Complete the attached fax form and fax it to 1-833-254-2597 or email it to [email protected].
- Quantity in commerce
- 149
- Distribution
- Distributed to accounts in AL AZ CA CT FL HI IA IL IN KY LA MA MD ME MI MN MO NC NH NJ NY OH OR PA PR SC TN TX VA WA WI
- Product codes and lots
- All units may experience this issue.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1249-2019 | Class II | Terminated | Voluntary: Firm initiated |