Z-1785-2019Class II
Liquid Cardiac Control, Model Nos. CQ5051, CQ5052, CQ5053
Randox Laboratories, Limited / initiated 2019-03-11 / Terminated
Randox Laboratories Ltd. began a recall of Liquid Cardiac Control, Model Nos. CQ5051, CQ5052, CQ5053 on . FDA classified it as Class II and gives the reason as "Customers may observe a decrease in recovered concentration with regards to NTproBNP in the Liquid Cardiac Control lots identified, on the Siemens Dimension EXL LOCI". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1785-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- Liquid Cardiac Control, Model Nos. CQ5051, CQ5052, CQ5053
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Customers may observe a decrease in recovered concentration with regards to NTproBNP in the Liquid Cardiac Control lots identified, on the Siemens Dimension EXL LOCI.
- FDA root cause
- Under Investigation by firm
- Action
- On March 11, 2019, the firm notified affected consignees of the product issue through an Urgent Medical Device Correction letter. Customers were informed that they may observe a decrease in recovered concentration for NTproBNP in the referenced Liquid Cardiac Control lots compared to the quoted target value on the value sheet for this analyzer only (Siemens Dimension EXL LOCI). Customers were asked to do the following: * Inspect your stock and quarantine affected stock. " Replace the value sheet in the kit with the revised value sheet provided. " Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. " Discuss the contents of this notice with your Medical Director. " Complete the response form even if you no longer have the affected product. Return the response form to [email protected] within five working days.
- Quantity in commerce
- 11420 kits
- Distribution
- Product distributed to CA, DE, WV, and Puerto Rico.
- Product codes and lots
- lots 4243CK, 4244CK, 4245CK.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1785-2019 | Class II | Terminated | Voluntary: Firm initiated |