Z-1247-2019Class II

SOMATOM Definition AS with Option syngo DE Scan for Single Source (Dual Spiral Dual Energy scan mode), Model...

Siemens Medical Solutions USA, Inc. / initiated 2019-02-14 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Definition AS with Option syngo DE Scan for Single Source (Dual Spiral Dual Energy scan mode), Model Number 8098027 Product Usage: Computed tomography... on . FDA classified it as Class II and gives the reason as "Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1247-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Definition AS with Option syngo DE Scan for Single Source (Dual Spiral Dual Energy scan mode), Model Number 8098027 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube.
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued letter dated 2/14/19 providing reason for recall and a user workaround in the Customer Safety Advisory Notice (CSAN) CT006/19/S. The recall issue will be permanently resolved by software update syngo CT VB10A_SP1. Contact the service organization 1-800-888-7436.
Quantity in commerce
53 total units distributed in U.S.
Distribution
US Nationwide Distribution
Product codes and lots
Serial numbers of all affected models: 122022 100143 100107 83834 83746 83811 66649 122020 83801 122008 122029 83767 83852 100092 83798 100091 100112 100129 100136 83837 122023 83871 100117 100097 100099 122009 83763 83819 83752 83845 83846 83749 83847 100146 83786 83842 83775 100090 83796 83768 83859 100087 83826 83754 83790 122011 83824 100147 83810 83758 100140 100085 83808

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1247-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1244-2019Class IITerminatedVoluntary: Firm initiated
Z-1245-2019Class IITerminatedVoluntary: Firm initiated
Z-1246-2019Class IITerminatedVoluntary: Firm initiated
Z-1247-2019Class IITerminatedVoluntary: Firm initiated