Medivance(R) ARCTIC SUN(R) ARCTICGEL(TM) Pads: Description, Catalog Number a) Intrepid ArcticGel Small Pad...
Bard Medical Division began a recall of Medivance(R) ARCTIC SUN(R) ARCTICGEL(TM) Pads: Description, Catalog Number a) Intrepid ArcticGel Small Pad Kit, 31705I b) Intrepid ArcticGel Medium Pad Kit... on . FDA classified it as Class II and gives the reason as "Specific lots of ArcticGel Pads have potential issues which may result in the insufficient adhesion of the hydrogel to the pad". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1241-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- C.R. Bard Inc
- Product
- Medivance(R) ARCTIC SUN(R) ARCTICGEL(TM) Pads: Description, Catalog Number a) Intrepid ArcticGel Small Pad Kit, 31705I b) Intrepid ArcticGel Medium Pad Kit, 31707I c) Intrepid ArcticGel Large Pad Kit, 31709I d) Intrepid ArcticGel Universal Pad, 31700I e) ArcticGel Universal Pad, 31700 f) ArcticGel Universal Pad, 4-pack, 3170004 g) ArcticGel X-Small Pad, 31703 h) ArcticGel X-Small Pad, 2-pack, 3170302 i) ArcticGel Small Pad Kit, 31705 j) ArcticGel Small Pad Kit, 2-pack, 3170502 k) ArcticGel Medium Pad Kit, 31707 l) ArcticGel Medium Pad Kit, 2-pack, 3170702 m) ArcticGel Large Pad Kit, 31709 n) ArcticGel Large Pad Kit, 2-pack, 3170902 o) ArcticGel Small Universal Pad Kit, 31810 p) ArcticGel Small Universal Pad,4-pack, 3180104 q) ArcticGel Neonatal Pad Kit, 31802 (international distribution only) r) Intrepid ArcticGel Small Pad Kit, 31705I Product Usage: The Arctic Sun Temperature Management System is a thermal regulating system, indicated for monitoring and controlling patient temperature. ArcticGel Pads are used with the Arctic Sun Temperature Management System to provide a hydrogel layer that ensures contact between the patient s skin and the pad throughout the treatment.
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- Specific lots of ArcticGel Pads have potential issues which may result in the insufficient adhesion of the hydrogel to the pad.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm sent a Voluntary Medical Device Product Notification by mail on 11/08/2017. The letter identified the affected product, problem and actions to be taken. The letter provided instruction on the proper way to remove the pad should the hydrogel peels away from the pad.
- Quantity in commerce
- 7526 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Argentina, Australia, Canada, Saudi Arabia, South Korea, Singapore, Thailand, and Chile
- Product codes and lots
- Catalog Number / lot number a) Intrepid ArcticGel Small Pad Kit, 31705I / NGAXY602 b) Intrepid ArcticGel Medium Pad Kit, 31707I / NGAXY603 c) Intrepid ArcticGel Large Pad Kit, 31709I / NGAXY604 d) Intrepid ArcticGel Universal Pad, 31700I / NGAXY601 e) ArcticGel Universal Pad, 31700 / NGAW3898, NGAW3055, NGAZ3417, NGBP1302, NGBP2293, NGBQ1040 f) ArcticGel Universal Pad, 4-pack, 3170004 / NGAW2916, NGAW3893, NGAZ3735, NGBP2294, NGBQ1021 g) ArcticGel X-Small Pad, 31703 / h) ArcticGel X-Small Pad, 2-pack, 3170302 / NGBP0392, NGBN3519 i) ArcticGel Small Pad Kit, 31705 / NGAW3894, NGAW4510, NGAZ2312, NGAZ3312, NGBN2492, NGBP0513 j) ArcticGel Small Pad Kit, 2-pack, 3170502 / NGAW3900, NGAW3059, NGAY4735, NGAZ0554, NGBP0514, NGBP4515 k) ArcticGel Medium Pad Kit, 31707 / NGBN3536, NGBP0512 l) ArcticGel Medium Pad Kit, 2-pack, 3170702 / NGBN3521 m) ArcticGel Large Pad Kit, 31709 / n) ArcticGel Large Pad Kit, 2-pack, 3170902 / o) ArcticGel Small Universal Pad Kit, 31810 / NGBP1534 p) ArcticGel Small Universal Pad,4-pack, 3180104 / NGBP1533, NGBP2320 q) ArcticGel Neonatal Pad Kit, 31802 (international distribution only) / NGBP3712 r) ArcticGel Neonatal Pad, 2-pack, 3180202 / NGBP0393