Z-1184-2019Class II
Diego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XS
Gyrus ACMI, Inc. / initiated 2019-03-08 / Terminated / root cause: software
Gyrus ACMI, Inc. began a recall of Diego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XS on . FDA classified it as Class II and gives the reason as "Olympus Diego Elite Consoles may inadvertently permit activation of the RF energy feature when used with Diego Elite blades". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1184-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gyrus ACMI, Inc.
- Product
- Diego Elite Turbinate Blades Bipolar Blade, 4mm curved, Product Number: BB4040XS
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Olympus Diego Elite Consoles may inadvertently permit activation of the RF energy feature when used with Diego Elite blades
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Olympus issued a voluntary field action on March 8, 2019 via letter and reply form sent to all consignees informing them of the risk and requiring them to take action as follows: c. Consignees in the United States using MDCONSlOO consoles with the affected software wiIl be instructed to cease using the consoles. Replacement consoles that do not utilize the affected software will be made available on an interim basis until updated software is made available, and can be loaded into the affected consoles by Olympus Field Service personnel. d. OUS Consignees instructed to remove the affected blades currently in their possession from use, and returned to Olympus. Replacement blades will be sent to these consignees as soon as the process flow changes to the respective production lines are implemented, and inspections of blade inventory in Olympus' possession are completed. e. All affected consoles will be loaded with updated software as soon as it is available.
- Quantity in commerce
- 193
- Distribution
- Nationwide Foreign: Germany,Japan, Singapore
- Product codes and lots
- All lot numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1178-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1179-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1180-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1181-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1182-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1183-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1184-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1185-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1186-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1187-2019 | Class II | Terminated | Voluntary: Firm initiated |