Z-1137-2019Class II

Biomerieux Vitek 2 Test kit VITEK 2 is an automated system consisting of instruments, software and reagent...

bioMerieux, Inc. / initiated 2019-03-15 / Terminated / root cause: software

bioMerieux, Inc. began a recall of Biomerieux Vitek 2 Test kit VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and... on . FDA classified it as Class II and gives the reason as "Unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1137-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
Biomerieux Vitek 2 Test kit VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
Unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits.
Root cause tag
software
FDA root cause
Software change control
Action
On 03/15/2019, the firm, BioMerieux sent notifications to all their consignees with the following instructions: "Please take the following actions at this time: Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. Apply the VITEK 2 Systems bioART rule criteria as previously mentioned (i.e., using the criteria listed below for ESBL / E. coli isolates). Refer to Appendix A for detailed instructions. o If the card is AST-NXXX (where XXX indicates the card type designation, 254 / 330 / 371) o The organism is Escherichia coli o And the phenotype is ESBL o And the drug cefuroxime is <=8 o And the drug ceftazidime is <=0.5 o Then stop for consultation and add a comment: There is no evidence that the isolate is an ESBL producer. The category results for cephalosporins should be interpreted as tested, before the AES expertise. Please store this letter with your bioMrieux instrument documentation. Complete the Acknowledgement Form and return it to your local bioMrieux representative. bioMrieux, Inc. is committed to providing our customers with the highest quality products, and we apologize for any inconvenience this may have caused in your laboratory. If you have any questions or concerns, please contact your local bioMrieux representative."
Quantity in commerce
10784 cartons
Distribution
Internationally to: United Kingdom, Latvia, Poland, Ukraine, Germany, and Scotland.
Product codes and lots
Ref: 413723 (AST-N254), 418674 (AST-N330) and 422024 (AST-N371 Card) with 8.01/9.01 SW  Unique Device Identifier: 03573026404550 (AST-N254) / 03573026474836 (AST-N330) / 03573026604363 (AST-N371)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1137-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1137-2019Class IITerminatedVoluntary: Firm initiated