POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE isp, 8F...
Bard Peripheral Vascular Inc began a recall of POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE isp, 8F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA... on . FDA classified it as Class II and gives the reason as "Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with...". As of , the recall status on record is Completed, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2080-2019
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE isp, 8F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 6F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 8F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; NAUTILUS DELTA Tip Confirmation System (Includes: Netbook with Pre-loaded Software, Patient Module, and ECG Cable) Product Usage: Is an implantable access device designed to provide repeated access to the vascular system.
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with the NAUTILUS DELTA" TCS.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On June 25, 2018 Bard sent recall notifications to all their customers informing them of the following: "This letter is to inform you of a voluntary product recall issued by Bard Peripheral Vascular, Inc. (BPV), a wholly owned subsidiary of Becton, Dickinson and Company. Specific product code / lot number combinations of the NAUTILUS DELTA" Port Kits are affected as outlined in Attachment 3. These product code / lot number combinations cannot be used with the NAUTILUS DELTA" Tip Confirmation System (TCS), product code 9770323. All other product code / lot number combinations not listed in Attachment 3 can continue to be used by your facility as they are safe to use and are not affected by this product recall. Reason for Recall: Per the current Instructions for Use (IFU), the NAUTILUS DELTA" TCS can be used with implantable ports; however, BPV has identified that the NAUTILUS DELTA" Port Kit product code / lot number combinations listed in Attachment 3 may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with the NAUTILUS DELTA" TCS. During port placement procedures, once the guidewire and dilator have been removed and the catheter is inserted into the introducer sheath, the catheter is flushed with saline through the port flushing hub with attached NAUTILUS DELTA" E Electrical Adaptor (provided with the NAUTILUS DELTA" Port Kits). This should cause an intravascular ECG waveform to appear; however, in multiple instances, it has been reported that no intravascular waveform is generated or the intravascular waveform is generated and subsequently lost. Clinical Risk Statement: The loss of or lack of intravascular signal while advancing the port catheter eliminates the devices intended use, which is to provide catheter tip navigation and positioning guidance based on specified ECG signals. This lack of consistent intravascular signal may lead to a prolonged procedure or inabi
- Quantity in commerce
- 95 units
- Distribution
- US Nationwide Distribution in the states of FL, CO, WI, OH, IL, KS, AZ, WV, CT, PA, AL, MI and VA
- Product codes and lots
- REBR1147, REBR1148, REBR1149, REBT0471, RECQ2162 and REBR1150
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2080-2019 | Class II | Completed | Voluntary: Firm initiated |