Innova IGS 6, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images...
GE Healthcare, LLC began a recall of Innova IGS 6, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and... on . FDA classified it as Class II and gives the reason as "There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1201-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Innova IGS 6, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customer were informed that they can continue to use their Innova IGS 6 or Innova IGS 630 Systems but must follow the following safety instructions: 1. Do not change the size of the field of view from 30cm to 20cm or 20cm to 30cm while releasing the fluoroscopy footswitch pedal. 2. Only perform these 2 actions sequentially. 3. If the system becomes locked up, perform a reset to recover functionalities. Customers were also informed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to perform the correction. After the GE representative has updated the system, please destroy previous Innova IGS 6 SW package installation media related to this system (versions prior to IGS6_2.2.1 or IGS6_3.1.1). Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 22 devices
- Distribution
- Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.
- Product codes and lots
- Model Number 5771387; System ID (Serial Number, UDI Number): 501257IGS630 (B3-18-014, 01008406821246141118070021B3-18-014); 951486RU630 (B3-18-009, 01008406821246141118040021B3-18-009); 714993IGS630 (B3-18-004, 01008406821246141118020021B3-18-004); 906225EP1 (B3-18-015, 01008406821246141118070021B3-18-015); 906225CATH4 (B3-18-016, 01008406821246141118080021B3-18-016); 843797IGS30 (B3-18-030, 01008406821246141119010021B3-18-030); 4603984XV41 (B3-18-019, 01008406821246141118090021B3-18-019); MCARIGS630 (B3-18-008, 01008406821246141118040021B3-18-008); 509473RM1B (B3-18-028, 01008406821246141119010021B3-18-028); 414649IGS8 (B3-18-020, 01008406821246141118090021B3-18-020); HU1099VA07 (B3-18-024, 01008406821246141118110021B3-18-024); A670060221 (B3-19-002, 01008406821246141119020021B3-19-002); A5163326 (B3-18-007, 01008406821246141118030021B3-18-007); YV4504 (B3-18-018, 01008406821246141118090021B3-18-018); YV4505 (B3-18-027, 01008406821246141118120021B3-18-027); YV4507 (B3-18-022, 01008406821246141118110021B3-18-022); 0850060762 (B3-18-025, 01008406821246141118110021B3-18-025); 0850060763 (B3-18-026, 01008406821246141118120021B3-18-026); KW1155VA02 (B3-18-017, 01008406821246141118080021B3-18-017); PL1689VA01 (B3-18-021, 01008406821246141118100021B3-18-021); PL0302VA02 (B3-18-011, 01008406821246141118060021B3-18-011); 06189VAS11 (B3-18-029, 01008406821246141119010021B3-18-029)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1199-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1200-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1201-2019 | Class II | Ongoing | Voluntary: Firm initiated |