syngo.via syngo.CT Cardiac Planning, Model Number 10496180
Siemens Medical Solutions USA, Inc began a recall of syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software... on . FDA classified it as Class II and gives the reason as "There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1113-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of any such finding.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Urgent Customer Safety Advisory Notice letters dated 3/20/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-888-7436.
- Quantity in commerce
- 112 total
- Distribution
- US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
- Product codes and lots
- Software versions VB20A or VB30A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1112-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1113-2019 | Class II | Terminated | Voluntary: Firm initiated |