Z-1233-2019Class II

Siemens ADVIA Centaur aTG (100 test kit) Siemens Material Number (SMN): 10492398

Siemens Healthcare Diagnostics Inc. / initiated 2018-12-07 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of Siemens ADVIA Centaur aTG (100 test kit) Siemens Material Number (SMN): 10492398 on . FDA classified it as Class II and gives the reason as "Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1233-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens ADVIA Centaur aTG (100 test kit) Siemens Material Number (SMN): 10492398
Product code
JZO System, Test, Thyroid Autoantibody
Reason for recall
Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens issued Urgent Medical Device Corrections, to Siemens Healthcare Diagnostics customers in the United States on December 7, 2018 delivered to customers on December 10, 2018. Urgent Field Safety Notices sent out to the country organizations on December 7, 2018 for distribution to customers outside the United States. These letters inform customers that Siemens Healthcare Diagnostics identified a bias with the affected lots of ADVIA Centaur aTG and Atellica IM aTG and action to take and resolution of the issue with new lots of the reagents. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Quantity in commerce
10673 kits
Distribution
Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana
Product codes and lots
Lot Number/Exp. Date: 68748302 12/13/2018 68749302 12/13/2018 83577304 1/12/2019 88637304 1/12/2019 5610306 2/11/2019 96436306 2/11/2019 19097308 3/14/2019 31866310 4/27/2019 45386310 4/27/2019 55627312 5/27/2019 69208316 6/29/2019 73000316 6/29/2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1233-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1233-2019Class IITerminatedVoluntary: Firm initiated
Z-1234-2019Class IITerminatedVoluntary: Firm initiated
Z-1235-2019Class IITerminatedVoluntary: Firm initiated
Z-1236-2019Class IITerminatedVoluntary: Firm initiated