Z-1297-2019Class II

Elekta Unity systems

Elekta Limited / initiated 2019-04-03 / Terminated / root cause: labeling and UDI

Elekta Limited began a recall of Elekta Unity systems Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation... on . FDA classified it as Class II and gives the reason as "There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1297-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Elekta Unity systems Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Elekta notified customers on about 04/03/2019 via email with "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter. Instructions included to ensure all staff working with Elekta Unity are aware of the issue and recall information and post the recall notice in a place accessible to all users, e.g. Instructions for use, until the recall is closed. Additionally customer were asked to complete and return the Acknowledgement Form.
Quantity in commerce
9 devices
Distribution
Worldwide Distribution - US Nationwide in the states of Texas and Wisconsin. Distribution internationally to Denmark, England, Germany, Italy, Netherlands
Product codes and lots
Product Code: UNITY SYSTEM/13553-006/600003, UNITY SYSTEM/10817-011/600010, UNITY SYSTEM/10562-011/600016; UNITY SYSTEM/10719-T03001/600013, UNITY SYSTEM/30004070-001/600023, UNITY SYSTEM/10420-AVL-U/600008, UNITY SYSTEM/11014-UMCU-U/600007, UNITY SYSTEM/11611-40/600011, UNITY SYSTEM/11489-21/600009;   UDI: 1536549, 1536549

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1297-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1297-2019Class IITerminatedVoluntary: Firm initiated