Elekta Unity systems
Elekta Limited began a recall of Elekta Unity systems Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation... on . FDA classified it as Class II and gives the reason as "There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1297-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Unity systems Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Elekta notified customers on about 04/03/2019 via email with "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter. Instructions included to ensure all staff working with Elekta Unity are aware of the issue and recall information and post the recall notice in a place accessible to all users, e.g. Instructions for use, until the recall is closed. Additionally customer were asked to complete and return the Acknowledgement Form.
- Quantity in commerce
- 9 devices
- Distribution
- Worldwide Distribution - US Nationwide in the states of Texas and Wisconsin. Distribution internationally to Denmark, England, Germany, Italy, Netherlands
- Product codes and lots
- Product Code: UNITY SYSTEM/13553-006/600003, UNITY SYSTEM/10817-011/600010, UNITY SYSTEM/10562-011/600016; UNITY SYSTEM/10719-T03001/600013, UNITY SYSTEM/30004070-001/600023, UNITY SYSTEM/10420-AVL-U/600008, UNITY SYSTEM/11014-UMCU-U/600007, UNITY SYSTEM/11611-40/600011, UNITY SYSTEM/11489-21/600009; UDI: 1536549, 1536549
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1297-2019 | Class II | Terminated | Voluntary: Firm initiated |