Z-1682-2019Class II
RUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0
Teleflex Medical / initiated 2019-04-01 / Terminated / root cause: manufacturing process
Teleflex Medical began a recall of RUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0 on . FDA classified it as Class II and gives the reason as "The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1682-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- RUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0
- Product code
- BTR Tube, Tracheal (W/Wo Connector)
- Reason for recall
- The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated the recall by letter on 04/01/2019. The letter stated reason for the recall and requested the return of the recalled units. Distributors were directed to collect the recalled units from their customers, consolidate them, and complete and return the acknowledgement form. The recalled units will be returned.
- Quantity in commerce
- 1787 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 14151, 14161, 14171, 14211, 14231, 14271, 14281, 14321, 14441, 14461, 14501, 14511, 15031, 15051, 15091, 15101, 15251, 15271, 15411, 16051, 16071, 16101, 16161, 16171, 16201, 16231, 16271, 16291, 16301, 16311, 16361, 16381, 16401, 16421, 16431, 16451, 16481, 16501, 16511, 17021, 17031, 17051, 17071, 17091, 17111, 17121, 17161, 17181, 17191, 17201, 17251, 17261, 17271, 17281, 17351, 17371, 17461, 18021, 18031, 18061, 18101, 18121, 18161, 18201, 18351, 18361, 18401
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1682-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1683-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1684-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1685-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1686-2019 | Class II | Terminated | Voluntary: Firm initiated |