Z-1721-2019Class II
Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution
Sybron Dental Specialties, Inc. / initiated 2018-07-13 / Terminated / root cause: labeling and UDI
Kerr/Pentron, DBA Kerr Corporation and Pentron Clinical began a recall of Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution on . FDA classified it as Class II and gives the reason as "Immersion time stated on the bottle is 45 minutes while the product instructions say 90 minutes". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1721-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sybron Dental Specialties, Inc.
- Product
- Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution
- Product code
- MED Sterilant, Medical Devices
- Reason for recall
- Immersion time stated on the bottle is 45 minutes while the product instructions say 90 minutes.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On July 13, 2018, Metrex Research Corp., a subsidiary of Kerr Corporation, sent URGENT NOTICE: Regimen Glutaraldehyde letters to their consignees instructing them to check their inventory and complete the enclosed Field Action Acknowledement and Product Return Form. Distributors were requested to conduct sub-recalls to their consignees.
- Quantity in commerce
- 31904
- Distribution
- AZ, CA, CO, FL, GA, IL, KY, LA, MA, MN, MO, NJ, NY, OH, PA, SC, TN, TX, UT, VA, and WA.
- Product codes and lots
- Part Number: 10-6465 (McKesson Reorder # 341) Lot Numbers: 16 1250, 16 1274, 16 1324A, 16 1324B, 16 2337, 17 1038, 17 1038A, 17 1083, 17 1193, 17 1262, 17 2122, 17 2142, 17 2219, 17 2306, 17 2347A, 18 1092B, 18 1121, 18 2037, and 18 3164.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1721-2019 | Class II | Terminated | Voluntary: Firm initiated |