VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW
Ortho-Clinical Diagnostics began a recall of VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600... on . FDA classified it as Class II and gives the reason as "Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1364-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Ortho Clinical Diagnostics issued on 20 March 2019, a customer letter via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) customers who were shipped affected VITROS XT 7600 Systems. Software Version 3.4.2 (MOD 4) will be provided for download or DVD format. Contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
- Quantity in commerce
- 85 units: US=19;OUS=66
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines
- Product codes and lots
- Serial Number 76000108 thru current.