Discovery NM/CT 670 ES
GE Healthcare, LLC began a recall of Discovery NM/CT 670 ES Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid... on . FDA classified it as Class II and gives the reason as "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1260-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Discovery NM/CT 670 ES Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 3 units
- Distribution
- US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
- Product codes and lots
- Mfg. Lot or Serial # System ID Model, Catalogue or Code # ESDX35020 403777D670C H3100BL ESPX54001 A51773134 H3100BM ESPY54002 DE203284 H3100BM
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1257-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1258-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1259-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1260-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1261-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1262-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1263-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1264-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1265-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1266-2019 | Class II | Terminated | Voluntary: Firm initiated |