Z-1497-2019Class II

Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine...

Abbott Medical / initiated 2019-04-05 / Terminated / root cause: design

Abbott Gmbh & Co. KG began a recall of Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER... on . FDA classified it as Class II and gives the reason as "Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1497-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.
Root cause tag
design
FDA root cause
Under Investigation by firm
Action
Abbott notified customers on about 04/05/2019 via "Product Correction Urgent - Immediate Action Required" letter. The letter instructed customers to power off their Alinity i System if they experience unexpected reagent supply temperature errors, detect a burning smell or visible smoke. They were also instructed that their Abbott representative will contact them to schedule the replacement of the impacted parts, to forward the Product Correction letter if the device has been further distributed, retain the letter for laboratory records, and complete and return the attached Customer Reply form. Questions can be directed to Customer Service at at 18774ABBOTT (available 24 hours a day, 7 days a week) for US customers. For customers outside the US, they are instructed to contact their local area Customer Service.
Quantity in commerce
373 units total
Distribution
Worldwide Distribution - US Nationwide to Oklahoma and Texas. Distributed worldwide to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, MADAGASCAR, MEXICO, NETHERLANDS, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UAE, UNITED KINGDOM, VIETNAM.
Product codes and lots
Part Number: 03R65 01, all Serial Numbers;   a. TED (Thermo Electric Device) Engine, Reagent cooler - Part Numbers A 30103732 01, A 30103732 02 with all Serial Numbers;   b. CABLE, POWER OUT, TED CNTRLER - Part Numbers A 35006203 01, A 35006203 02 with all Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1497-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1497-2019Class IITerminatedVoluntary: Firm initiated