Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200
Siemens Medical Solutions USA, Inc began a recall of Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray... on . FDA classified it as Class II and gives the reason as "Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1295-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued Customer Safety Advisory Letter dated 3/22/19 advising of problem, health risk and action to take: recommends to avoid overwriting the organ programs, as this can lead to frequent operations. Siemens is distributing this letter and the addendum to the user manual to all potentially affected users Contact our service organization for any additional questions at 1-800-888-7436.
- Quantity in commerce
- 2 US
- Distribution
- US Nationwide Distribution in the states of NC, MA
- Product codes and lots
- Cios Alpha system with SW VA30
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1295-2019 | Class II | Terminated | Voluntary: Firm initiated |