Siemens Medical Solutions USA, Inc. / initiated 2019-03-22 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray... on . FDA classified it as Class II and gives the reason as "Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1295-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued Customer Safety Advisory Letter dated 3/22/19 advising of problem, health risk and action to take: recommends to avoid overwriting the organ programs, as this can lead to frequent operations. Siemens is distributing this letter and the addendum to the user manual to all potentially affected users Contact our service organization for any additional questions at 1-800-888-7436.
Quantity in commerce
2 US
Distribution
US Nationwide Distribution in the states of NC, MA
Product codes and lots
Cios Alpha system with SW VA30

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1295-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1295-2019Class IITerminatedVoluntary: Firm initiated