Z-1927-2019Class III

cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval

Roche Diagnostics GmbH / initiated 2019-04-04 / Terminated / root cause: manufacturing process

Roche Diagnostics Corporation began a recall of cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval on . FDA classified it as Class III and gives the reason as "There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1927-2019
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On April 4, 2019, Roche issued Urgent Medical Device Correction notices (UMDC TP-00620) to customers via UPS Ground (receipt required). Customers were advised to take the following actions: 1) Inspect the tray input flap on the cobas p 501 or p 701 post-analytical unit to determine if it is loose. If it is not fastened securely or becomes loose prior to the part being replaced, contact your Roche Field Engineer Specialist (FES) to repair the instrument. 2) Complete and return the fax form (TP-00621) via fax or email 3) File the UMDC for future reference. Customers with questions may contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in the UMDC.
Quantity in commerce
4 units
Distribution
US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA
Product codes and lots
Serial Numbers - 1001 through 1044

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1927-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1927-2019Class IIITerminatedVoluntary: Firm initiated
Z-1928-2019Class IIITerminatedVoluntary: Firm initiated