Z-1318-2019Class II

SureSigns VS3 NBP Temp, Product Number 863070. For monitoring the physiological parameters of patients (NBP...

Philips North America LLC / initiated 2019-04-19 / Open, Classified / root cause: software

Philips North America, LLC began a recall of SureSigns VS3 NBP Temp, Product Number 863070. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature) on . FDA classified it as Class II and gives the reason as "System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1318-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
SureSigns VS3 NBP Temp, Product Number 863070. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Root cause tag
software
FDA root cause
Software design
Action
On April 19, 2019 Philips Healthcare issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. Actions to be taken by customer: Upon receipt of this notification, promptly perform the software upgrade to each of your SureSigns VS3/VS4 monitors, as called out in the attached Pulse Rate Software-Installation Instructions. It details how to obtain and install the system software update. It also provides instructions on how to export the file that is created during software installation and instructions where to email the file back to Philips. If you need any further information or support concerning this issue, please contact your local Philips representative or call us at 1-800-722-9377.
Quantity in commerce
518
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.
Product codes and lots
Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1318-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1317-2019Class IIOngoingVoluntary: Firm initiated
Z-1318-2019Class IIOngoingVoluntary: Firm initiated
Z-1319-2019Class IIOngoingVoluntary: Firm initiated
Z-1320-2019Class IIOngoingVoluntary: Firm initiated
Z-1321-2019Class IIOngoingVoluntary: Firm initiated
Z-1322-2019Class IIOngoingVoluntary: Firm initiated
Z-1323-2019Class IIOngoingVoluntary: Firm initiated
Z-1324-2019Class IIOngoingVoluntary: Firm initiated