Z-1387-2019Class II

DEXIS Titanium Intraoral Sensor

KaVo Dental Technologies, LLC / initiated 2019-04-29 / Terminated / root cause: design

KaVo Dental Technologies LLC began a recall of DEXIS Titanium Intraoral Sensor Product Usage: The DEXIS Titanium Intraoral Sensor is a USB-driven sensor which is intended to acquire dental intraoral... on . FDA classified it as Class II and gives the reason as "The sensor's look up calibration files were incorrectly programmed during manufacturing, which can result in the reduction of image quality". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1387-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
KaVo Dental Technologies, LLC
Product
DEXIS Titanium Intraoral Sensor Product Usage: The DEXIS Titanium Intraoral Sensor is a USB-driven sensor which is intended to acquire dental intraoral radiographic images. The DEXIS / KaVo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intra-oral radiographs. The DEXIS / KaVo sensor can be used either in combination with special positioning devices to facilitate positioning and alignment with the x-ray beam or it may also be positioned by hand with the assistance of the patient.
Product code
MUH System, X-Ray, Extraoral Source, Digital
Reason for recall
The sensor's look up calibration files were incorrectly programmed during manufacturing, which can result in the reduction of image quality.
Root cause tag
design
FDA root cause
Device Design
Action
KaVo Dental Technologies, LLC notified customers on about 04/29/2019, via "URGENT: MEDICAL DEVICE RECALL - CORRECTION" letter. Customers were instructed to: 1.) Review the serial number information included with this letter and identify where the affected devices are located in your facility; 2.) For any affected devices that are installed, stop usage and contact Customer Care at 1-888-883-3947 (Option 1) to start the correction via remote connection. See Supplemental Directions on Page 2; 3.) For any affected devices that are unopened or uninstalled, return them to KaVo Dental Technologies, LLC. See Supplemental Directions on Page 2; 4.) If you have sold or forwarded the affected devices to a location outside your control, see Supplemental Directions on Page 2; 5.) Your response is required: Complete and sign the bottom portion of this letter. Return a copy of completed and signed Page 1 to confirm your Recall Responsibility is complete. Note: This can be completed by dental office personnel or on your behalf by your Authorized Dealer. Correction via Reprogramming by a Customer Care Representative: 1.) If you would like to complete your Recall Responsibility by a correction of any affected devices that are already installed, this can be completed free of charge by a Customer Care Representative via remote connection. 2.) The issue can be resolved through a simple and brief software reprogramming activity by replacing the calibration correction file through a remote connection. The sensor hardware meets all specifications, and following the software reprogramming activity, the sensor will operate as intended. 3.) To begin, please contact Customer Care at 1-888-883-3947 (Option 1). 4.) The Customer Care Representative will provide instructions on how to complete the software correction via remote connection. 5.) However, if you prefer to return your affected devices that are installed instead of remote connection, follow the return direction
Quantity in commerce
211 devices
Distribution
Worldwide Distribution - US Nationwide to AL, AR, AZ, CA, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NV, OH, OK, PA, TN, TX, VA, WA, WI, WV and internationally to Canada.
Product codes and lots
Model Number CD63M115.CG; Part Number IOXS-1.012.6180; Serial Numbers: 13095594, 13095595, 13095599, 13095661, 13095665, 13095759,  13095762, 13095766, 13095769, 13095774, 13095780, 13095782, 13095783, 13095784, 13095786, 13095809, 13095810, 13095817, 13095823, 13095827, 13095839, 13095876, 13095890, 13095897, 13095899, 13095900, 13095921, 13095927, 13095929, 13095941, 13095942, 13095955, 13095957, 13096033, 13096034, 13096037, 13096038, 13096040, 13096063, 13096064, 13096065, 13096120, 13096128, 13096145, 13096151, 13096155, 13096162, 13096173, 13096193, 13096195, 13096200, 13096206, 13096212, 13096217, 13096218, 13096220, 13096223, 13096224, 13096227, 13096230, 13096234, 13096235, 13096244, 13096252, 13096257, 13096267, 13096272, 13096286, 13096290, 13096292, 13096295, 13096299, 13096311, 13096318, 13096322, 13096335, 13096340, 13096344, 13096346, 13096348, 13096355, 13096362, 13096383, 13096391, 13096402, 13096405, 13096410, 13096417, 13096426, 13096428, 13096430, 13096437, 13096439, 13096466, 13096479, 13096483, 13096490,  13096491, 13096504, 13096505, 13096518, 13096527, 13096564, 13096574, 13096576, 13096580, 13096583, 13096591, 13096593, 13096602, 13096610, 13096622, 13096725, 13096738, 13096760, 13096804, 13096825, 13096831, 13096883, 13096905, 13096916, 13096928, 13096947, 13096958, 13096960, 13096964, 13096968, 13096970, 13096975, 13096979, 13096982, 13096984, 13096993, 13097017, 13097018, 13097038, 13097039, 13097101, 13097109, 13097118, 13097121, 13097126, 13097153, 13097166, 13097170, 13097181, 13097190, 13097193, 13097195, 13097199, 13097211, 13097214, 13097217, 13097234, 13097240, 13097241, 13097247, 13097249, 13097252, 13097269, 13097287, 13097299, 13097306, 13097309, 13097320, 13097322, 13097326, 13097348, 13097386, 13097389, 13097398, 13097418, 13097419, 13097422, 13097438, 13097439, 13097464, 13097469, 13097476, 13097478, 13097503, 13097517, 13097518, 13097532, 13097546, 13097553, 13097554, 13097557,  13097559, 13097562, 13097580, 13097584, 13097589, 13097603, 13097615, 13097616, 13097619, 13097808, 13097827, 13097838, 13095874, 13095917,  13095930, 13096057, 13096240, 13096448, 13097145, 13097230, 13097245, 13097463, 13097813.

Enforcement reports