IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 -
Philips North America, LLC began a recall of IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 - Product Usage: Used primarily as a traditional telemetry... on . FDA classified it as Class II and gives the reason as "The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1692-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.
- Product code
- DSI Detector And Alarm, Arrhythmia
- Reason for recall
- The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On March 28, 2019, Philips Healthcare distributed URGENT MEDICAL DEVICE RECALL notices (FSN 86201863A) dated March 22, 2019 to customers via courier service. Actions To Be Taken by Customer/User at the time was as follows: - MX4O monitors can continue to be used while awaiting a software upgrade. - Customers should maintain an adequate supply of fully charged batteries given the approximately 25% increased power consumption. Switch to continuous or manual SO2 measurements if your unit is using automatic measurement mode. On April 24, 2019, Philips Healthcare distributed UPDATE URGENT MEDICAL DEVICE RECALL notices (FSN86201863B) dated April 16, 2019 to customers via courier service. Actions To Be Taken by Customer/User is currently as follows: Until you have upgraded the software on your MX4Os, replace your batteries every 8 hours, unless you are using continuous monitoring mode and Sp02 monitoring with AA batteries, in which case replacement should be done every 2 hours. Philips Healthcare will release a software upgrade, which will correct this problem. Philips will contact you to arrange for software correction. If you need any further information or support concerning this issue please contact your local Philips representative or Philips Customer Support at 1-800- SUPPORT 722-9377.
- Distribution
- Worldwide distribution - US Nationwide and to 71 countries.
- Product codes and lots
- All revision B.X versions are affected.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1692-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1693-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1694-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1695-2019 | Class II | Ongoing | Voluntary: Firm initiated |