Boston Scientific Corporation began a recall of AURIGA 30 GENERAL SYSTEM UPN: M0068S30G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system... on . FDA classified it as Class II and gives the reason as "Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1397-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- AURIGA 30 GENERAL SYSTEM UPN: M0068S30G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Boston Scientific issued to the user level the correction/removal via a customer notification letter via FedEx or hand delivery by a Boston Scientific agent on 4/9/19. Letter provides recommendations to the user until a field repair can be completed or a replacement console can be provided. A Boston Scientific representative will be in contact with you directly to schedule a field repair or replacement at your facility. The console can continue to be used during this time following the recommendations within the letter.
- Quantity in commerce
- 18 (OUS)
- Distribution
- Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
- Product codes and lots
- Serial Number: 21302793, 21302835, 21302799, 21302844, 21302861, 21302780, 21302921, 21302999, 21303007, 21303013, 21303012, 21303064, 21303032, 21303033, 21303045, 21303065, 21302816, 21302994
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1391-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1392-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1393-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1394-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1395-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1396-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1397-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1398-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1399-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.