TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China)...
Abbott Laboratories Inc. (St Jude Medical) began a recall of TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP) on . FDA classified it as Class II and gives the reason as "In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1493-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)
- Product code
- OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Reason for recall
- In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recall notice, and clinician acknowledgment form were hand delivered by Abbott Sales Representatives to all impacted accounts starting April 5, 2019. Completed Clinician Acknowledgement Forms, signify they received the correction notification for the impacted TactiSys Quartz Equipment with software version 1.7.
- Quantity in commerce
- 317 units
- Distribution
- US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA
- Product codes and lots
- UDI: 07640157990194 Serial Numbers: 33019, 33059, 33350, 33354, 33357, 33360, 33361, 33363, 33390, 33392, 33393, 33405, 33417, 33421, 33444, 33449, 33451, 33457, 33458, 33461, 33491, 33492, 33509, 33013, 33015, 33020, 33040, 33075, 33091, 33097, 33160, 33197, 33384, 33385, 33387, 33389, 33414, 33433, 33388, 33435, 33481, 33154, 33157, 33162, 33163, 33278, 33092, 33159, 33437, 33412, 33014, 33027, 33028, 33029, 33038, 33049, 33057, 33082, 33096, 33192, 33222, 33268, 33379, 33383, 33397, 33398, 33399, 33400, 33422, 33423, 33478, 33320, 33352, 33395, 33351, 33069, 33381, 33455, 33064, 33093, 33313, 33337, 33161, 33426, 33083, 33321, 33420, 33051, 33155, 33196, 33256, 33353, 33460, 33217, 33427, 33073, 33150, 33076, 33148, 33434, 33396, 33050, 33333, 33333, 33336, 33336, 33095, 33143, 33143, 33112, 33113, 33266, 33267, 33269, 33270, 33271, 33272, 33273, 33274, 33275, 33276, 33283, 33287, 33288, 33291, 33325, 33356, 33359, 33362, 33404, 33459, 33016, 33098, 33099, 33100, 33101, 33102, 33103, 33104, 33105, 33106, 33107, 33108, 33110, 33111, 33127, 33130, 33131, 33265, 33018, 33024, 33025, 33087, 33152, 33156, 33220, 33230, 33231, 33254, 33279, 33281, 33315, 33416, 33428, 33438, 33439, 33442, 33443, 33463, 33330, 33386, 33228, 33033, 33048, 33149, 33259, 33282, 33464, 33022, 33031, 33035, 33037, 33044, 33045, 33046, 33055, 33066, 33067, 33068, 33078, 33080, 33085, 33086, 33088, 33094, 33129, 33137, 33138, 33140, 33144, 33158, 33195, 33218, 33221, 33229, 33253, 33257, 33262, 33280, 33289, 33293, 33326, 33328, 33355, 33364, 33382, 33403, 33406, 33408, 33409, 33411, 33424, 33431, 33440, 33445, 33446, 33448, 33450, 33454, 33462, 33465, 33466, 33169, 33011, 33012, 33021, 33023, 33034, 33054, 33056, 33074, 33079, 33089, 33090, 33109, 33191, 33193, 33194, 33215, 33224, 33314, 33316, 33327, 33329, 33358, 33375, 33376, 33407, 33415, 33419, 33429, 33447, 33219, 33258, 33292, 33334, 33335, 33032, 33410, 33081, 33170, 33171, 33172, 33173, 33198, 33199, 33201, 33203, 33204, 33205, 33206, 33207, 33208, 33209, 33210, 33211, 33212, 33213, 33214, 33294, 33295, 33296, 33298, 33299, 33300, 33301, 33302, 33303, 33304, 33365, 33366, 33367, 33368, 33369, 33370, 33371, 33372, 33373, 33374, 33549, 33550, 33551, 33168, 33165, 33216, 33042, 33456, 33070, 33432, 33480, 33047, 33072, 33077, 33136, 33226, 33263, 33377, 33378,