Artis Zee (material # 10094135, 10094137, 10094141) -
Siemens Medical Solutions USA, Inc began a recall of Artis Zee (material # 10094135, 10094137, 10094141) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for... on . FDA classified it as Class II and gives the reason as "In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1339-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis Zee (material # 10094135, 10094137, 10094141) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee and Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On April 9, 2019, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that ARTIS systems equipped with certain motor control units may encounter with hardware failure, and planned procedures may need to be terminated and performed on an alternative x-ray system. The firm stated their service organization would contact customers shortly to arrange a date to perform a software update to correct the issue. For an earlier appointment or if you have any questions, you may contact the firm at 1-800-888-7436. Customers were asked to promptly notify and instruct all staff at the organization who need to be aware of the notice. Additionally, if you further distributed the product, please identify your customers and notify them at once of this recall.
- Quantity in commerce
- 459 (116 US)
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL,AR, AZ, CA, FL, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NV, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, VA , WA WI and WV. and Foreign countries.
- Product codes and lots
- Serial Numbers 138260 138221 138259 138250 138268 138266 124286 125245 148402 148401 148384 148390 147249 148393 148411 148400 148358 148361 148397 148392 148382 148363 148375 155237 155203 155226 155227 155235 131235
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1339-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1340-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1341-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1342-2019 | Class II | Terminated | Voluntary: Firm initiated |