Z-1687-2019Class II

Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Abbott Vascular / initiated 2019-05-01 / Terminated

Abbott Vascular began a recall of Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only on . FDA classified it as Class II and gives the reason as "Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1687-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Vascular
Product
Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
Product code
NKM Mitral Valve Repair Devices
Reason for recall
Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.
FDA root cause
Use error
Action
On 05/01/19, Urgent Medical Device Correction notices were sent to implanting physicians, via personal delivery or email, informing them that under normal use conditions, this failure will not occur. To prevent unintended force from being applied to the clip, revised instructions, including definition and related technique, for performing the steps "Establish Final Angle" and "Invert the Clip Arms" were developed and provided herein. The revised instructions will be applied to the shared XTR and NTR Clip Delivery System Instructions for Use. Unintended excessive force applied to the NTR Clip Delivery System during "Establish Final Arm Angle" can result in unexpected movement of clip arms; however, unlike XTR, the NTR Clip does not become damaged and remains functional. Adhering to the revised IFU sections for both product platforms will ensure consistency of use. Further, the firm requested that customers read the Field Safety Notice, review revised instructions for use with firm representatives, sign and return Effectiveness Check Forms, and share this information with all personnel associated with the procedure. The revised instruction steps provided in this communication will be updated in the product IFU and included in associated training. Your current inventory of product is acceptable for safe use following the revised IFU steps described above. There is no need to return any product. Customers with additional questions are encouraged to call customer service at 800-227-9902.
Distribution
U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM
Product codes and lots
All Serial Numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1687-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1687-2019Class IITerminatedVoluntary: Firm initiated