NeuViz 16 Essence Multi-slice CT Scanner System
Neusoft Medical Systems Co., Ltd. began a recall of NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system... on . FDA classified it as Class II and gives the reason as "Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1281-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Neusoft Medical Systems Co., Ltd.
- Product
- NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A customer notification letter will be sent to affected customers by email. All the affected systems will be corrected with the upgraded software in the field free of charge.
- Quantity in commerce
- 2
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- Series numbers N16E180037E; N16E180047E
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1279-2019 | Class II | Terminated | FDA Mandated | ||
| Z-1280-2019 | Class II | Terminated | FDA Mandated | ||
| Z-1281-2019 | Class II | Terminated | FDA Mandated |