Opteform Allograft Disc, 90mm x 5mm, 22cc, Catalog Number 600-03-90
Exactech, Inc. began a recall of Opteform Allograft Disc, 90mm x 5mm, 22cc, Catalog Number 600-03-90 on . FDA classified it as Class II and gives the reason as "The Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1690-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Exactech, Inc.
- Product
- Opteform Allograft Disc, 90mm x 5mm, 22cc, Catalog Number 600-03-90
- Product code
- MQV Filler, Bone Void, Calcium Compound
- Reason for recall
- The Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- The firm initiated the recall on 04/26/2019 by electronic mail. The notice requested the consignee cease distribution of the product, notify their customers (the user), quarantine all units subject to recall in inventory, and if any units were implanted to follow-up on the patient's condition 30 days after surgery.
- Quantity in commerce
- 3 units
- Distribution
- OK, VA, FL
- Product codes and lots
- UDI 10885862094100 Serial Numbers: T31977961, T31977962, T31977968
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1688-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1689-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1690-2019 | Class II | Terminated | Voluntary: Firm initiated |