Z-1934-2019Class II

VITROS 250AT Chemistry System, Product Code 1758143, (UDI # 10758750000036)

Ortho-Clinical Diagnostics, Inc. / initiated 2019-04-08 / Terminated / root cause: design

Ortho-Clinical Diagnostics began a recall of VITROS 250AT Chemistry System, Product Code 1758143, (UDI # 10758750000036) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral... on . FDA classified it as Class II and gives the reason as "Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1934-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 250AT Chemistry System, Product Code 1758143, (UDI # 10758750000036) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
Root cause tag
design
FDA root cause
Device Design
Action
On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions: - As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits. - If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides. - If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use. - Consider loading one cartridge of VITROS CK Slides on board your system at a time. - Complete and return the request form via fax to 1-888-557-3759. Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take.
Quantity in commerce
57 units
Distribution
US Nationwide Distribution
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1934-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1933-2019Class IITerminatedVoluntary: Firm initiated
Z-1934-2019Class IITerminatedVoluntary: Firm initiated
Z-1935-2019Class IITerminatedVoluntary: Firm initiated
Z-1936-2019Class IITerminatedVoluntary: Firm initiated