Z-2133-2019Class II

Nasal Alar SpO2 Sensor, Adult/Pediatric SpO2 Sensor, Rx Only

Invivo Corporation / initiated 2019-05-08 / Terminated / root cause: design

Xhale Assurance, Inc. began a recall of Nasal Alar SpO2 Sensor, Adult/Pediatric SpO2 Sensor, Rx Only Product Usage: The Assurance Nasal Alar SpO2 Sensor is indicated for single patient use for... on . FDA classified it as Class II and gives the reason as "These Instructions for Use has missing instructions related to the checking and changing of the application site procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2133-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Invivo Corporation
Product
Nasal Alar SpO2 Sensor, Adult/Pediatric SpO2 Sensor, Rx Only Product Usage: The Assurance Nasal Alar SpO2 Sensor is indicated for single patient use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients, (at least 4 years and older and weighing > 15kg). The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision
Product code
DQA Oximeter
Reason for recall
These Instructions for Use has missing instructions related to the checking and changing of the application site procedure.
Root cause tag
design
FDA root cause
Device Design
Action
Xhale Assurance, Inc. notified consignees on about 05/08/2019, via "URGENT - Medical Device Correction" letter. Customers were instructed to insert the IFU Addendum into each case of sensors for ready reference, review all information provided in the updated IFU Addendum with all members of your staff whom need to be aware of the updated instructions, and complete and return the provided Customer Reply Form. Distributors were instructed to insert the IFU Addendum into each case of sensors for ready reference, provide a copy of the Field Safety Notice and IFU Addendum to any customer to whom affected devices have been distributed to, and complete and return the provided Customer Reply Form.
Quantity in commerce
701,272 devices total
Distribution
US Nationwide distribution to AZ, CA, FL, KS, LA, MA , MO, MS, NC, NY, OH, PA, TX, WI. International distribution to Canada, Germany, Malaysia, United Kingdom.
Product codes and lots
Reference Codes (Lot Numbers): 0303 (101619, 101624, 101628, 101632, 101637, 101642, 101646, 101647, 101651, 101701, 201651, 101707, 101715, XH101711, XH101720, XH101715, XH101726, XH101729, XH101733, XH101737, XH101820, XH101842), 301-11214 (TR101711, TR101715, TR101720, TR101726, TR101729, TR101744, TR101733, TR101737, TR101748, TR101742, TR101746, TR101750, TR101803, TR101811, TR101817, TR101820, TR101824, TR101829, TR101833, TR101838, TR101842, TR101847, TR101852), 0201-A01(PF101736, PF101806, PF101814, PF101817, PF101819, PF101822, PF101828, PF101833, PF101838). Instructions for Use for revisions 10412_7 and prior (i.e. 10412_6, 10412_5, etc) as well as 10358_6 and prior.  Shelf life is 3 years for sensors manufactured prior to Oct 2018. 5 years for sensor manufactured on or after Oct 2018.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2133-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2133-2019Class IITerminatedVoluntary: Firm initiated