LEEP Precision Integrated System 120V, Model LP-10-120
CooperSurgical, Inc. began a recall of LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION... on . FDA classified it as Class II and gives the reason as "2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1840-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperSurgical, Inc.
- Product
- LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .
- Product code
- HGI Electrocautery, Gynecologic (And Accessories)
- Reason for recall
- 2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On April 23, 2019, the firm distributed Urgent Medical Device Safety Alert letters to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to immediately discontinue use of the discrepant product, quarantine the product, and replace the 2 main fuses with the 2 fuses provided with the firm's mailing. Customers were provided with instructions to replace the fuses. Once customers have successfully completed the fuse replacement, they are asked to complete the Acknowledgement and Receipt form and return it to CooperSurgical. If you have any questions, please contact the firm at 203-601-5200 ext. 3300.
- Quantity in commerce
- 19
- Distribution
- US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.
- Product codes and lots
- 2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,2018110013, 2018110014, 2018110015,2018110016, 2018110017, 2018110019, 2018110020,2018110021,& 2018110022