Z-1937-2019Class II
iChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is...
Siemens Healthcare Diagnostics Inc. / initiated 2018-10-18 / Terminated
Beckman Coulter Inc. began a recall of iChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland... on . FDA classified it as Class II and gives the reason as "Falsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1937-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- iChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland, Made in Germany.
- Product code
- KQO Automated Urinalysis System
- Reason for recall
- Falsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction.
- FDA root cause
- Component change control
- Action
- The recalling firm issued letters dated 10/15/2018 on 10/18/2018 via email or regular mail for customers where an email was unavailable. The letter explains the issue and impact, and provides the actions the consignee is to take when using the product. The letter informed the customer they are actively working on a resolution to prevent recurrence of the issue.
- Quantity in commerce
- 956,798/100 test strip vials
- Distribution
- Distribution was nationwide, including Puerto Rico. There was also military/government distribution. Foreign distribution was made to Canada, Argentina, Brazil, Chile, Colombia, Mexico, Peru, Philippines, Singapore, and Taiwan.
- Product codes and lots
- Lot numbers beginning with lot 7212143 except for: 7212146, 7212148, 7212160, 7212162, 7212163 and 7212164.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1937-2019 | Class II | Terminated | Voluntary: Firm initiated |