ETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX
Ethicon, Inc. began a recall of ETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in... on . FDA classified it as Class II and gives the reason as "Possibility that a suture raw material containing high endotoxin levels was used in the manufacturing of this lot". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1842-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon, Inc
- Product
- ETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures
- Product code
- GAQ Suture, Nonabsorbable, Steel, Monofilament And Multifilament, Sterile
- Reason for recall
- Possibility that a suture raw material containing high endotoxin levels was used in the manufacturing of this lot
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Ethicon notified primary and secondary US consignees by telephone on 4/19/19 and follow-up letter dated April 24, 2019 stating reason for recall, health risk and action to take: examine inventory, remove product for return. Complete the Business Reply Form (BRF) to Stericycle at 1-844-721-3045 or [email protected]. Questions regarding this action or to report any customer complaints, please contact your Ethicon sales representative or contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 8:00 AM to 6:00 PM ET.
- Quantity in commerce
- 36 eaches
- Distribution
- Worldwide - US Nationwide Distribution - IA, NJ, RI, TX, WV Foreign: Canada
- Product codes and lots
- Lot Code:PBQ797
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1841-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1842-2019 | Class II | Terminated | Voluntary: Firm initiated |