Z-1906-2019Class II

EVOLIS Microplate System (Catalog no. 89700) with TimeLiner Data Management Software, (Catalog no. 90471A

Bio-Rad Laboratories Inc. / initiated 2019-01-10 / Open, Classified / root cause: software

Bio-Rad Laboratories, Inc began a recall of EVOLIS Microplate System (Catalog no. 89700) with TimeLiner Data Management Software, (Catalog no. 90471A on . FDA classified it as Class II and gives the reason as "User-induced circumstances can contribute to the EVOLIS Microplate Processor System not handling repeat testing requests/results properly and may contribute to transmitting erroneous results to the customer's Laboratory...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1906-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Bio-Rad Laboratories Inc.
Product
EVOLIS Microplate System (Catalog no. 89700) with TimeLiner Data Management Software, (Catalog no. 90471A.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
User-induced circumstances can contribute to the EVOLIS Microplate Processor System not handling repeat testing requests/results properly and may contribute to transmitting erroneous results to the customer's Laboratory Information System (LIS).
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued letters dated 1/9/2019 via FedEx on 1/10/2019 to all EVOLIS customers.
Quantity in commerce
213 devices
Distribution
Distribution was nationwide to medical facilities, including to Puerto Rico. There was also government/military distribution. There was no foreign distribution.
Product codes and lots
Software version 1.4, UDI for the software is 03610520562024.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1906-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1906-2019Class IIOngoingVoluntary: Firm initiated