Z-2066-2019Class I

Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2019-05-03 / Terminated / root cause: manufacturing process

Medline Industries Inc began a recall of Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the... on . FDA classified it as Class I and gives the reason as "The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2066-2019
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the subject of this recall, and the recalled BD Alaris Pump Infusion Set (RES#82951) Product Usage: The Vyaire enFlow Disposable Cartridges contain a heating element that is used to help raise or maintain a patient's body temperature.
Product code
N/A
Reason for recall
The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm notified its distributor, Medline, of the recall via Urgent Product Recall Notice. The distributor was asked to identify all inventory within its possession and to cease distribution immediately. The distributor was also asked to forward a copy of the notice to any customers to whom the product may have been further distributed. Medline notified customers in a letter dated March 20, 2019, titled "VYAIRE MEDICAL SUB-RECALL IMMEDIATE ACTION REQUIRED". Customers were instructed to do the following: 1. Immediately check your stock for the specific kit number and the affected kit lot number (attached list). Quarantine any kits with the listed lot numbers. If you dont have any of the lots listed above skip to step 3. 2. Affix the enclosed stickers on the affected kits and return to stock. If additional stickers are needed please call 866-359-1704. Please be sure NOT to use the recalled enFlow Disposable Cartridge components, pull another approved product from sterile supply. 3. Please fill out and return the enclosed Verification Form via email to [email protected] or fax 866-767-1290. Customers with questions should contact the firm at 1-866-359-1704.
Quantity in commerce
4,680 kits
Distribution
One distributor located in Illinois.
Product codes and lots
Lots 2018090690 2018101190 2018101590 2018121190 2019013190 2019030490 2019040190

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2066-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2066-2019Class ITerminatedVoluntary: Firm initiated