Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 20L GP Bath Model Number: FSGPD20
Thermo Fisher Scientific (Asheville) LLC began a recall of Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 20L GP Bath Model Number: FSGPD20 on . FDA classified it as Class II and gives the reason as "Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1867-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Thermo Fisher Scientific (Asheville) LLC
- Product
- Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 20L GP Bath Model Number: FSGPD20
- Product code
- JTQ Bath, Incubators/Water, All
- Reason for recall
- Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Thermo Fisher issued on 4/02/19 via Email notification with attachments to distributors (domestic and foreign) the Urgent Medical Recall Notification Letter with customer response sheet. Recall is to the user level.On May 23, 2019 email followed by a formal written, "Urgent Medical Device Recall Notification Letter" with customer response sheet were sent via US mail to domestic customers. Letters address health risk and action to take: customer service team is prepared to send you a replacement immediately and the necessary materials and instructions for the return of the bath you received initially to the factory. Please call representatives with your original purchase order, company name, and contact information at 866-984-3766 or email the included Recall Customer Response Form.
- Quantity in commerce
- 1 unit
- Distribution
- CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
- Product codes and lots
- Serial Numbers: 300247790
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1865-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1866-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1867-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1868-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1869-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1870-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1871-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1872-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1873-2019 | Class II | Terminated | Voluntary: Firm initiated |