Hudson RCI Sheridan LTS, Product Code 5-11112
Teleflex Medical began a recall of Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management on . FDA classified it as Class I and gives the reason as "This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1759-2019
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
- Product code
- BTR Tube, Tracheal (W/Wo Connector)
- Reason for recall
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111
- Quantity in commerce
- 3483 units
- Distribution
- Worldwide - US Nationwide Distribution
- Product codes and lots
- Lot Numbers: 73K1600369, 73L1600011, 73A1700646, 73C1700238, 73D1700625, 73E1700475, 73F1700387, 73H1700041, 73H1700483, 73J1700292, 73L1700503, 73M1700453, 73A1800671, 73B1800350, 73D1800138, 73F1800174, 73G1800126, 73G1800806, 73J1800308, 73K1800173, 73L1800225, 73K1800172
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1759-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1760-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1761-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1762-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1763-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1764-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1765-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1766-2019 | Class I | Terminated | Voluntary: Firm initiated |