Egia 60 Curved Tip Art Med Thick Sulu, Egia60ctamt
Covidien LLC began a recall of EGIA 60 CURVED TIP ART MED THICK SULU, EGIA60CTAMT on . FDA classified it as Class II and gives the reason as "The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1830-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Egia 60 Curved Tip Art Med Thick Sulu, Egia60ctamt
- Product code
- GDW Staple, Implantable
- Reason for recall
- The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology. " Complete the Recalled Product Return Form. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Quantity in commerce
- 3,113,280
- Distribution
- Nationwide domestic distribution, worldwide foreign distribution.
- Product codes and lots
- N8L0311KY N8M0063KY
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1825-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1826-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1827-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1828-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1829-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1830-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1831-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1832-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1833-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1834-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1835-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1836-2019 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.