Signia Tri-Staple 2.0 30Mm Medium/ Thick Reload , Sig30amt
Covidien LLC began a recall of SIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMT on . FDA classified it as Class II and gives the reason as "The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1832-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Signia Tri-Staple 2.0 30Mm Medium/ Thick Reload , Sig30amt
- Product code
- GDW Staple, Implantable
- Reason for recall
- The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology. " Complete the Recalled Product Return Form. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Quantity in commerce
- 3,113,280
- Distribution
- Nationwide domestic distribution, worldwide foreign distribution.
- Product codes and lots
- N6E0033UX N6H0768UX N6H0772UX N6K0521UX N6K0522UX N6K0667UX N6K0668UX N6L0687UX N6M0399UX N7A0018UX N7A0213UX N7A0401UX N7A0837UX N7B0787UX N7C0238X N7C0470X N7E0496X N7E1088X N7F0670X N7H0962X N7J0370X N8A0188X N8B0057X N8B0483X N8B0940X N8D0365X N8D0649X N8E0092X N8E0213X N8E0516X N8E1178X N8F0483X N8F0819X N8F0974X N8J0440X N8L0589Y N8M0529Y N9A0050Y N9A0427Y
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1825-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1826-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1827-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1828-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1829-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1830-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1831-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1832-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1833-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1834-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1835-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1836-2019 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.