Z-1889-2019Class II
Ethicon ENSEAL X1 Curved Jaw Tissue Sealer 37cm Shaft
Ethicon Endo-Surgery Inc / initiated 2019-05-15 / Terminated
Ethicon Endo-Surgery Inc began a recall of Ethicon ENSEAL X1 Curved Jaw Tissue Sealer 37cm Shaft PRODUCT CODE: NSLX137C on . FDA classified it as Class II and gives the reason as "Observed reported incidence rate for harm associated with intraoperative bleeding". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1889-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- Ethicon ENSEAL X1 Curved Jaw Tissue Sealer 37cm Shaft PRODUCT CODE: NSLX137C
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Observed reported incidence rate for harm associated with intraoperative bleeding
- FDA root cause
- Under Investigation by firm
- Action
- Ethicon issued Urgent Medical Device recall letter dated May 15, 2019 advising of product problem, health risk, and action to take: determine if you have product subject to this recall (removal) on hand and quarantine such product(s); 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall (removal) letter when communicating. 4. Complete the Business Reply Form (BRF) 5. Keep this notice visibly posted for awareness until all product subject to this recall (removal) has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. 6. Customers are required to return unused impacted ENSEAL X1 Curved Jaw Tissue Sealer in their inventory immediately. Questions customer complaints contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 7:30 AM to 6:30 PM ET.
- Quantity in commerce
- 1059 boxes
- Distribution
- Nationwide Foreign: CANADA, AUSTRALIA, BELGIUM
- Product codes and lots
- All lot numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1888-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1889-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1890-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.