Z-2201-2019Class II

The Alinity ci-series System Control Modules which are configured with Alinity c Processing Modules. The...

Abbott Medical / initiated 2019-05-21 / Terminated / root cause: software

Abbott Gmbh & Co. KG began a recall of The Alinity ci-series System Control Modules which are configured with Alinity c Processing Modules. The Alinity ci-series Control Module is labeled in part,"*... on . FDA classified it as Class II and gives the reason as "All versions of the Alinity ci-series software may not detect an issue on the Alinity c Integrated Chip Technology (ICT) assays, sodium (Na+), potassium (K+), and chloride (Cl-)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2201-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
The Alinity ci-series System Control Modules which are configured with Alinity c Processing Modules. The Alinity ci-series Control Module is labeled in part,"* * *Alinity ci-series* * *SYSTEM CONTROL MODULE Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. The Alinity c Processing Module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c Processing Module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c Processing Module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
All versions of the Alinity ci-series software may not detect an issue on the Alinity c Integrated Chip Technology (ICT) assays, sodium (Na+), potassium (K+), and chloride (Cl-).
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
On May 21, 2019, the firm sent Product Correction Urgent - Immediate Action Required letters to all of their consignees which advised them of the problem and provided them with instructions for preforming additional testing. US customers were sent this letter and an attached Customer Reply Immediate Action Required form vis FedEx priority overnight express. Abbott service personnel will follow-up to install new software.
Quantity in commerce
943 systems
Distribution
Worldwide Distribution - US Nationwide AL, AR, CA,, FL, GA, ID, IL, LA, MA, ME, MI, MN, MO, NJ, NY, OK, SC, TN, TX, UT, VA, WA, WI, and PUERTO RICO. ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM,BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLUMBIA, CROTIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDOESIA, IRELAND, ISRAEL, JORDAN, KENYA, KUWAT, LATVIA, LEBONON, MEXICO, NAMIBA, NETHERLANDS, NORWAY, PACKISTAN, PHILLINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDON, and VIETNAM.
Product codes and lots
List Number: 03R70-01 ALL LOTS

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2201-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2201-2019Class IITerminatedVoluntary: Firm initiated