Z-2135-2019Class II

Stratus CS Acute Care cTNI TestPak, SMN 10445071

Siemens Healthcare Diagnostics Inc. / initiated 2019-05-29 / Terminated

Siemens Healthcare Diagnostics Inc began a recall of Stratus CS Acute Care cTNI TestPak, SMN 10445071 Product Usage: In vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma... on . FDA classified it as Class II and gives the reason as "Potential for increased rate of false positive results and/or increased rate of random non-repeatable false positive results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2135-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Stratus CS Acute Care cTNI TestPak, SMN 10445071 Product Usage: In vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma. Cardiac troponin I measurements can be used as an aid in the diagnosis of acute myocardial infarction (AMI).
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
Potential for increased rate of false positive results and/or increased rate of random non-repeatable false positive results.
FDA root cause
Under Investigation by firm
Action
On May 29, 2019, Siemens issued Urgent Medical Device Correction notices to U.S.A. customers via FedEx courier service and Urgent Field Safety Notices to international customers via email. Actions to be Taken by the Customer: 1. Do not use TestPaks within 2 mos of expiration. 2. Repeat samples with cTNL results above 0.07 ng/mL or your institution's established 99th percentile value. 3. Complete and return the Field Correction Effectiveness Check form 4. Retain a copy of the letter with your laboratory records and forward a copy of the notice to anyone to whom you may have distributed the product. Siemens is working to resolve this issue and will contact customers when additional information is available. Siemens will reimburse users for repeat tests and discarded cTNL TestPaks within 2 months of the printed expiration date associated with this field corrective action.
Quantity in commerce
2312
Distribution
US Nationwide Distribution
Product codes and lots
Lot # 238330002 forward Expiration Dates - May 25, 2019 - November 2, 2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2135-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2135-2019Class IITerminatedVoluntary: Firm initiated