Z-1910-2019Class II

Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging...

GE Healthcare / initiated 2019-05-16 / Terminated / root cause: design

GE Healthcare, LLC began a recall of Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging sources, and from other healthcare information sources on . FDA classified it as Class II and gives the reason as "There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1910-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.
Root cause tag
design
FDA root cause
Device Design
Action
On May 20, 2019, the firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter date 5/20/2019 to their consignees. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: discontinue use of the UV study management functionality for study split/study info updates until a correction is available and to wait for a GE representative to contact them to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
145 (87 US 103 OUS)
Distribution
Worldwide Distribution: US (nationwide) to states of: FL, GA, IL, IN, LA, MA, MO, NJ, NY, OR, PA, SC, SD, TX, and WA; and countries of : Brazil, Canada, Chile, France, French Polynesia, Germany, India, Italy, Japan, Kuwait, Lebanon, Singapore, Spain and United Kingdom.
Product codes and lots
UDI # 840682103800  Model # 2098249-002 Software Version: 6.0.9, 6.0SP9.01, CPACS 6.0 SP9.0.1, UV 6.0 SP9.0.1, UV 6.0 SP9.01.1, UV 6.0 SP9.0.1.2, UV 6.0 SP9.01.3, UV6.0 SP9.01.4, UV 6.0 SP9.01.5, UV 6.0 SP9.0.1.6, UV Version 6.0 SP5.0.4.1, UV Version 6.0 SP9, UV Version 6.0 SP9.0.0.1, UV Version 6.0 SP901, UV Version 6.0 SP9011, and UV Version 6.0 SP902

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1910-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1910-2019Class IITerminatedVoluntary: Firm initiated
Z-1911-2019Class IITerminatedVoluntary: Firm initiated