Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2279-2019Class II

BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001

Becton Dickinson & Co / initiated 2019-05-30 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations on . FDA classified it as Class II and gives the reason as "Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2279-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations
Product code
N/A
Reason for recall
Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no longer than two days. If the sample was stored for more than 2 days, the result for blood alcohol determination might not be accurate (either falsely low or falsely high)
Root cause tag
manufacturing process
FDA root cause
Process control
Action
BD PAS notified consignees via UPS and email on May 30, 2019. A corrected letter was issued dated June 12, 2019, the sku and lot number for the one affected lot of product was correct, the product name was incorrect. The notice should replace the initially distributed notice dated May 30, 2019. The letter identifies reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific catalog and lot number listed above. Destroy all product subject to the recall in accordance with your institutions process for destruction. 2. Share this Urgent Medical Device Recall notification with all users of the product in your facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response/Certificate of Destruction Form and return to the BD contact noted on the form regardless of whether you have any affected material or not so that BD may acknowledge your receipt of this notification and process your product replacement, if applicable. Contact: Customer Quality 888-237-2762 OPT 3, OPT 2 Monday Friday 8:00am and 5:00pm (CT)
Quantity in commerce
247,000 tubes
Distribution
US Nationwide-Distribution in the states of CA, FL, GA, IL, KS, MO, PA, RI, TN Foreign: Canada
Product codes and lots
Lot Number: 8187663 Exp. Date: 2020/7/31   UDI: (01)30382903670018 (17)200731(10)8187663 (30)0100

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2279-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2279-2019Class IITerminatedVoluntary: Firm initiated