Siemens Artis zeego--Interventional Fluoroscopic X-Ray System Model Number: 10280959
Siemens Medical Solutions USA, Inc began a recall of Siemens Artis zeego--Interventional Fluoroscopic X-Ray System Model Number: 10280959 on . FDA classified it as Class II and gives the reason as "Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1897-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Artis zeego--Interventional Fluoroscopic X-Ray System Model Number: 10280959
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced.
- Quantity in commerce
- 29
- Distribution
- Nationwide
- Product codes and lots
- Serial Numbers: 160335 160430 160368 160469 160357 160370 160369 160509 160427 160433 160443 160428 160429 160463 160001 160461 160486 160800 160487 160488 160507 160840 160862 160856 160823 160914 160930 161235 160401
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1894-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1895-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1896-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1897-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1898-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1899-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1900-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1901-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1902-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1903-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1904-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1905-2019 | Class II | Terminated | Voluntary: Firm initiated |